4. Select “START/STOP”, then the LED on the key will light in green and the Pressure Inspection
will be performed.
Criteria : Make sure that the difference between the displayed pressure and the standard
pressure gauge is within the criteria.
Accuracy should be within ±3mmHg with the following measurement pressure 50,
100, 150, 200, 250, and 300mmHg.
If the accuracy does not meet the criteria, a malfunction of the pressure sensor
on the NIBP unit can be considered. Replace the NIBP unit.
9
-