Precautions for Installing the Equipment
This section describes the environmental condition to use the DSL-8001.
C A U T I O N
The installation of this equipment should be performed by our service
representative or a person who is well acquainted with this equipment.
Precautions about the Operating Environment
y
The following environmental conditions should be observed when operating the equipment.
・
Surrounding Temperature : 5 to 40
°
C
・
Relative Humidity
: 30 to 85% (non-condensing)
・
Atmospheric Pressure
: 700 to 1060hPa
y
The DSL-8100 is intended for patient monitoring in ICU, CCU, surgery, ER, and ward. Direct use
in MRI environment or home-care should be prohibited.
y
The power source should fulfill the following condition.
・
Verify power voltage and frequency before connecting to an AC power source.
・
Use the power source that can provide adequate power to the device.
y
Pay attention when installing or storing the equipment. Do not install or store in the following
locations.
・
where chemicals are stored or gas may evolved.
・
where the equipment will be subject to splashing water or humidity from a nebulizer or
vaporizer
・
where the equipment will be subject to direct sunlight
・
unstable place with inclination, vibration, or shock.
y
Ensure proper ventilation to cool the device.
・
A minimum space of 5 cm is required between vents on the rear side of the monitor and the
wall. If the monitor is embedded in a wall or surrounded by a wall, a minimum space of 10
cm is required.
y
This equipment must be placed on a level surface. If installed in incorrect direction, water or
chemicals may enter the equipment and cause damage. This may also cause paper jam in the
recorder unit.
W A R N I N G
If the equipment is used under an environment not fulfilling the specified
condition, not only that the equipment cannot deliver its maximum
performance, the equipment may be damaged and safety cannot be
ensured.
If using in an environment other than specified above, contact our service
representative.
C A U T I O N
Equipotential Grounding
When connecting multiple equipments, electrical potential difference may
be generated between the equipments. This may result in electric shock to
the patient connected to these equipments. Pay special attention for use in
Operating Room, ICU, CCU, Cardiac Catheter Laboratory, and
Cardiovascular X-ray room. To avoid such electrical potential difference,
use the ground cable to connect each equipment's potential equalization
terminal to the same ground terminal. This is called equipotential grounding
7
-
2
Summary of Contents for DSL-8001
Page 4: ...Blank Page ...
Page 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 44: ...2 6 Blank Page ...
Page 60: ...3 16 Blank Page ...
Page 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Page 64: ...4 4 Blank Page ...
Page 72: ...6 6 Blank Page ...
Page 102: ...7 30 Blank Page ...
Page 148: ...11 6 Blank Page ...