2
-
4
90bpm Bidirectional Systoles : 90bpm
Response time of heart rate meter to change in heart rate
HR change from 80bpm to 120bpm : Range 4.1 to 4.9 sec.
Average
4.4
sec.
HR change from 80bpm to 40bpm : Range 5.9 to 6.4 sec.
Average
6.1
sec.
Time to ALARM for tachycardia
Ventricular Tachycardia 1mVpp, 206bpm : Average 15.8 sec.
Ventricular Tachycardia 2mVpp, 206bpm
: Average 6.3 sec.
Ventricular Tachycardia 0.5mVpp, 206bpm : Average 13.3 sec.
Ventricular Tachycardia 2mVpp, 195bpm
: Average 7.8 sec.
Ventricular Tachycardia 4mVpp, 195bpm
: Average 8.4 sec.
Ventricular Tachycardia 1mVpp, 195bpm
: Average 7.9 sec.
Respiration
Method :
Impedance
Method
Frequency Characteristic : 2.5Hz
Current :
100
μ
A or lower
Measurement Range
: 0, 3 to 150Bpm
Measurement Accuracy
: ±3 Bpm
Temperature
Measurement Method
: Thermistor Method
Measurement Range
: 0 to 50°C / 32 to 122°F
Measurement Accuracy
: ± 0.2°C (probe and main unit)
± 0.1°C (main unit only)
SpO
2
(Arterial Oxygen Saturation)
Measurement Method
: 2 Wavelength Pulse Wave Method
Wavelength: Approx. 660nm (red light)
Approx. 890nm (infrared light)
Output: 15mW or below
Measurement Range
: 1 to 100%
Resolution :
1%
Measurement Accuracy
: Adult ± 2% at 70 to 100%
(± 3% when DS-100A is used)
Not specified when less than 70%.
Neonate ± 2% at 70 to 100%
Not specified when less than 70%.
PR Measurement Range
: 20 to 300bpm
PR Resolution
: 1bpm
PR Accuracy
: ± 3bpm for 20 to 250bpm
Not specified when 251 to 300bpm.
Measurement Update Rate
: 1 sec.
Measurement Response Time : 6 to 7sec. (averaging duration)
Summary of Contents for DSL-8001
Page 4: ...Blank Page ...
Page 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 44: ...2 6 Blank Page ...
Page 60: ...3 16 Blank Page ...
Page 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Page 64: ...4 4 Blank Page ...
Page 72: ...6 6 Blank Page ...
Page 102: ...7 30 Blank Page ...
Page 148: ...11 6 Blank Page ...