Equipotential Grounding
When connecting multiple equipments, electrical potential difference may be generated between
the equipments. This may result in electric shock to the patient connected to these equipments.
Pay special attention for use in Operating Room, ICU, CCU, Cardiac Catheter Laboratory, and
Cardiovascular X-ray room. To avoid such electrical potential difference, use the ground cable to
connect each device’s potential equalization terminal to the same ground terminal. This is called
equipotential grounding.
The ground cable is shaped like a connector, and
can be disconnected manually without using tools.