Replacement of the NIBP Unit
●
Usable Life of the NIBP Unit
The NIBP unit can measure approximately 100,000 times.
Replace the NIBP unit to maintain proper function since the air may leak and the unit may not
correctly function if the unit is used for a long time.
●
Tool
Phillips screwdriver (Large)
●
Disassembling/Assembling
Pay attention to wiring and tubing, and follow the procedure below to replace the NIBP unit.
1. Remove the 4 screws used to fix the NIBP
holder on the bottom of the main unit.
2. Hold the lock plate and remove the NIBP
holder and disconnect the air tube
(connected between the NIBP unit and the
main unit).
3. Remove the 2 screws and replace the unit with
a new NIBP unit.
4. Connect the NIBP unit to the air hose on the
main unit and insert the NIBP holder to the
main unit.
Air Tube
NIBP Unit
NIBP Holder
9
-