Questions and Troubleshooting
5.
Other Functions
(1) Method to execute [Dust Check] automatically after turning power on
• Refer to [Power On Settings] (Page 65).
(2) Method to hide messages on [READY] screen
• Change the fiber image from X/Y view to X magnified view or Y magnified view by
pressing
X/Y
.
(3) Too many repetitions until “Test Finish” indicated in [Arc Calibration]
• The splicer needs to repeat the arc calibration after replacing the electrodes or when the
environmental conditions change drastically.
(Page 65). The number of arc calibrations can be set to a specific amount. When the
splicer completes the set amount of calibrations, it indicates “Test Finish”. However,
this does not mean it is completely calibrated.
(4) “Test Finish” is never indicated after many repetitions in [Arc Calibration]
• Execute [Stabilizing Electrodes] function in [Maintenance Menu]. If the splicer still
does not indicate “Test Finish”, replace the electrodes. Refer to [Replace Electrode]
section.
(5) No arc power change after [Arc Calibration]
• An internal factor is calibrated and adjusted for the specific arc power selected. The
displayed arc power in each splice mode does not change.
• The calibration results affect all splice modes.
(6) Method to input different comments after each splice in splice results data
• Refer to [Splicing procedure] (Page 20).
(7) Method to input the same comments automatically in splice results data
• Refer to [Splicing procedure] (Page 20).
(8) Method to cancel storing splice results in internal memory
• Refer to [Splicing procedure] (Page 20).
(9) Method to download splice results from splicer to PC
• Contact your nearest sales agency.