31
Maintenance of Splicing Quality
2-2.
Rotating Cleaver Blade
If the cleaver does not cleave properly, rotate the blade 1/16th of a turn to replace the
worn out blade position with a sharp blade position. To rotate blade, do the following:
(1) Remove the blade cover.
(2) Using a small slotted screwdriver, loosen the blade lock screw.
(3) Rotate the circular blade 1/16th of a turn.
(4) Tighten the blade lock screw.
(5) Attach the blade cover.
Blade Height Adjustment
After the circular blade has been rotated a complete revolution (16 positions), blade
height needs to be adjusted to compensate for blade wear. Do the following:
(1) Loosen the blade lock screw.
(2) Using a .89mm hex wrench, loosen the adjuster lock screw.
(3) Using a 1.5mm hex wrench, turn the adjuster clockwise so that the reference dot
aligns with the next position mark. Never attempt to rotate the adjuster more than 2
position marks.
(4) Tighten the adjuster lock screw.
(5) Tighten the blade lock screw.
Blade Replacement
After the circular blade is raised 3 times, it needs to be replaced. Contact your Fujikura
splicer distributor.
• When rotating the blade, do not touch the cutting edge. Move the blade
with a thin cotton swab, for easy and safe rotation.
• After blade height adjustment, the 16 blade positions can be used again. If
the cleaver does not cleave properly, rotate the blade.
Blade lock screw
(Cone point set screw M3x4)
Adjuster lock screw
(Set screw M2x3)
Adjuster
initial
position
Direction of
rotation
Adjuster