53
Splice Menu
(4) The calibration result is displayed after completion.
“Test Finished” message
Arc power and splicing position
calibration are successfully completed.
Press to exit.
“Test Again” message
Arc power and splicing position
calibration are completed but further
calibration is strongly recommended,
as the change from the previous arc
calibration is too large.
Press to perform arc calibration, or
to exit even though arc calibration is
not completed.
• In some cases, multiple iterations of arc calibration are needed until the
calibration process is successfully completed and the "Test Finish" message is
displayed. Arc calibration can be considered almost completed if multiple
iterations are completed without receiving the message.
• Number threshold can be set so that "Test Finish" message is displayed after
specific number of arc calibrations are performed. See [Other Option] (Page
59)
for detail.
Arc calibration
Select the [Arc Calibration]
in the [Splice Menu]
Arc calibration
“
Finish
”