62
Management Menu
1.
Change of Operating Direction
The splicer is shipped from the factory with settings for the "Monitor Front" operation
style. This can be changed to "Monitor Rear" operation style. The sheath clamp direction
can also be changed for a selected style. When [Monitor Position] is changed, the
direction of the four arrow keys is reversed.
Changing monitor position
(1) Select a monitor position in [Monitor Position] menu.
(2) Turn off splicer and change monitor angle.
Monitor Front Monitor Rear
(3) Change direction of sheath clamps.
1. Open sheath clamp and loosen and remove screws with a 1.5 mm hexagonal
wrench, and remove sheath clamp cover to expose the shaft.
2. Remove the shaft. A paper clip is helpful to remove the shaft, and tweezers are
helpful for holding the shaft.
3. Take sheath clamps off.
4. Reverse direction of sheath clamp, insert shaft and tighten screws.
(4) Turn splicer Power on.