va06.0_0031g_Eng.fm
89
Inter
ve
n
tion p
roce
dur
es
Disconnect the flat cables connected to the CPU board at:
J2 and J13
J7, J3 and J10
J4 and J12
J1
Unscrew the fixing screws (marker 6) holding the CPU board to the flange.
J12
6
J1
J3
J2
J13
J4
J10
J7
Remove the CPU board unit being careful not to pull out the flat cable on the inside.
J14
J11
Disconnect the flat cables (J14 and J11) if necessary.