va06.0_0060g_Eng.fm
103
Fiche
s d’in
te
rven
tion
Unscrew the 2 screws (marker 6) of the air detector board (marker 7).
6
6
7
Remove the board being careful not to pull out the flat cable.
Disconnect the flat cable connected to the board at J1.
J1
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
Calibrate the air bubble detector.
Carry out the regular servicing tests (see Quality Control Certificate).