Operating mode
The software operates according to the operating mode described in the EN 60601-2-24
standard for
infusion pumps
.
Start the acquisition programme for the scale.
Enter the necessary programme data without confirming the
flow rate.
Adjust the device to the specific flow rate.
Confirm the flow rate on the micro-computer to enable self-
adjustment of the scale.
Start infusion by pressing "VALIDATION", when
00.00
appears on the scale display.
At the end of the specified time, note the error percentage
displayed on the screen.