40
va04.3_001g_Eng.fm
Ma
in
te
na
nc
e pr
év
en
tive
4.4.10 Test 8, battery information
This test displays:
the battery running time displayed in hours, days and months
the number of discharge cycles equivalent to the number of hours of battery
use / 7h
the start date for the battery running time counter; the battery voltage in mV
and in 1/100 volts
the battery temperature in mV and in 1/10 °C
the current charge/discharge in LSB and mA.
Test 8
, select "enter".
The "
Battery info
" main menu displays.
Choose between the "Clear"
and "exit" options that appear.
Select "Clear" to reset to 0 after all battery replacements.
"Clear", using the left selection key
The "
Battery Info
" menu displays.
Duration
:
the number of battery operating hours is reset to "0",
resetting the number of cycles to"0".
Cycles
:
the number of cycles is reset to "0".
Start date
:
Only the total battery running time since the beginning
remains incremented.
Scroll through the information, from screen to screen, using
the selection keys.
Conversion
:
Voltage
:
Voltage value at the battery terminals and conversion
in mV.
Conversion
:
Temperature
:
battery temperature value and conversion in mV.
Conversion
:
Current
:
the value of the current consumed or on charge and
its conversion in LSB.
Press "exit" to return to the "
Test
" menu.
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...