va04.3_001g_Eng.fm
Ma
in
te
na
nc
e pr
év
en
tive
4.4.13 Test 11, Ageing test
This text checks the motor block/pumping system is functioning correctly as well as the pump
occlusivity.
Test 11
, press "enter".
The main, "
Ageing test
" menu displays.
When the pump is fitted with a tube, open the
administration set clamp by pressing the bolus key.
Select your submenu using the "OK" key.
Flow rate
: between 0.1 and 1200ml/h
Modify the value using the following selection keys:
for the hundreds and tens, use the double arrows to
increase or decrease
for the units, increase or decrease using the single
arrows.
Press "OK" to validate and to move to the next step.
Mode
: Go forward or synchro or go forwards/backwards
or backwards
Modify your choice using the single arrow selection keys
.
Press "OK" to confirm and move to the next step.
Steps numbers
or turns
number
:
Content or value lying between 1 and 9999
Modify the value using the single arrow selection keys
Press "OK" to confirm and move to the next step.
Current
: high, low, auto or error.
Modify the value using the single arrow selection keys
,
Press "OK" to confirm and move to the next step.
Press "OK" to confirm and start up the test.
The "
Ageing test
" menu displays:
The flow rate and current category for the
motor command are displayed at the top
of the screen.
The number of steps decreases according to
the actual change for the number of steps.
Press "exit" to return to the "
Test
" menu.
During the test, the rotation control is activated and the device set in error if a problem is
detected.
In a test with a defined number of steps/cycles or using a synchro mode, the test ends
when the requested position is reached.
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...