4.5.4 SAV 3, Data log event
This test displays the list of the alarms stamped with date and time (like
Test 4
).
The "
Data log event
" main menu displays.
Scroll through the information, from screen to screen, using
the selection keys.
Press "enter" to display the alarm type.
Press "MENU/EXIT" to return to the "
Test
" menu.