58
va04.4_001e_Eng.fm
Pr
even
tive ma
inten
ance
4.6.2 Measuring with a scale
Use a new administration set [otherwise, make a hole in the filter of the administration set
(if equipped with) and use this set for maximum 10 calibrations].
Do not recycle medical water.
Keep the container clean changing it as often as necessary.
Make sure the environmental temperature is between 18°C and 30°C.
Equipment needed
Chronometer
Scale:
Flow rate value
Scale sensitivity
x < 5ml/h
1/10000th
5ml/h < x < 30ml/h
1/1000th
x > 30ml/h
1/100th
Test tube or beaker with 1ml graduations.
Liquid: distilled water, oil
Catheter extension with Luer Lock end (length 100cm, internal diameter 2.5mm)
Needle:
Flow rate value
Type of needle
x < 30ml/h
G26
x > 30ml/h
G18 or G21
Installation
Following the diagram below:
Anti-vibration measuring table
1 cm
Catheter extension bracket support
50 cm
Observe the equipment horizontal installation plan.
Fill the container with 600ml or more of distilled water.
Connect an administration set with the container.
Connect the canula with the distal end of the administration
set.
Fit the administration set on to the pump.
Fill the administration set with water. Take care that no air is
in the administration set before the measurement is started.
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...