124
va06.0_0100g_Eng.fm
F
ich
es d
’inter
vention
Dismantling
When working with electronic components, we recommend wearing an anti-static bracelet
connected to earth and working on a foam anti-static mat.
Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
Disconnect the flat cable connecting the CPU board to the power supply at J1.
Remove the RJ45 stopper (marker 3).
Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
4
2
J1
5
4
3
Remove the upper case (marker 5) being careful not to pull out the display board flat
cable.
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...