2.1.4 The mains board
The mains board contains the mains plug, the ECG filter as well as F1 and F2 fuses.
It is the interface between the mains and the power supply board. The mains board is situated
on the angle bracket. It enables the power supply board to be changed without the need for
soldering.
J1 on the inside
F2
F1
J2
Mains board
This board is connected to the other parts by means of connectors.
J1 connector to power supply board connector
Pin
Signal
Description
1
Neutral
Mains power supply
2
Phase
Mains power supply
3
GND
Ground
J2 connector to mains
Pin
Signal
Description
1
Neutral
Mains power supply
2
Phase
Mains power supply
3
GND
Functional earth
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...