![Fresenius Kabi Volumat Agilia Technical Manual Download Page 66](http://html.mh-extra.com/html/fresenius-kabi/volumat-agilia/volumat-agilia_technical-manual_2331623066.webp)
66
va05.1_001e_Eng.fm
T
rou
ble
sh
ootin
g
Flow rate
flashing alarm
Flow rate has
been modified but
has not been
confirmed
Check the flow rate and
confirm with the OK key
End of infusion
pre-alarm
The remaining
VTBI is less than
5% of the initially
programmed VTBI
or 5 minutes or
5 ml remain before
the initially
programmed VTBI
is reached
Check if the volume remaining
in the container corresponds to
the remaining VTBI. If necessary,
prepare a container for a new
infusion sequence. If a drop
detector is present, the end of
infusion pre-alarm is disactivated
and can be defined as an option.
End of infusion
alarm
VTBI delivered.
KVO activated
conforming with
the configuration
Press Stop to define the new
infusion parameters (if
necessary).
Drop detector
alarm (if used
with this
accessory)
The drop
detector is not
correctly
positioned
Check the drop detector
position
Check the integrity of the drop
detector
Check the drop detector
connection
Try the device using another
drop detector
If the trouble keeps up, check
the connection on the device
Replace the component if
necessary
The drop
detector has been
connected then
disconnected
The drop
detector is
disconnected
when the pump is
configured to
function with it
Connect the drop detector to
the pump and to the drip
chamber
Under flow rate
alarm
Flow rate
detected by the
drop detector is
lower than the set
flow rate
Check the container
Check the roller clamp
Check that liquid forms
properly approx. 20 drops / ml
Displayed
message
Meaning
Cause
Action
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...