2.2 Operational description
From an operational point of view, the
Volumat Agilia
pump
is made up of 3 sub-assemblies:
a sub-assembly to maintain and to check the correct positioning of the administration set.
a motorisation sub-assembly.
an external connection sub-assembly.
2.2.1 Administration set check and maintenance sub-assembly
The administration set is positioned along the front part panel held in place by the door.
The front part is equipped with four detection systems:
A detector to check the position of the door (closed /open).
A photoelectric cell to detect the presence of the administration set clamp.
An ultra-sound detector to detect the presence of air bubbles in the administration set.
Detection of the pressure is ensured by two piezo-resistant sensors allowing the pressure
level control in the administration set.
2.2.2 Pumping sub-assembly
The pumping sub-assembly contains the peristaltic pumping mechanism.
The mechanism is made up of a camshaft that creates a wave-like movement in a bank of 12
"fingers". The displacement of these ’fingers’ is controlled by the CPU board, while turning it
displaces the liquid at the programmed flow rate.
The camshaft is driven by a stepper-motor and a reduction gearbox assembly via a toothed
belt.
A photoelectric cell and a rotation palette, situated on the camshaft end, to control the motor/
pumping system rotation.
2.2.3 External connections sub-assembly
Volumat
Agilia
has 2 plugs situated at the back of the angle bracket:
a mains supply plug
a 8 point Binder plug which, together with various specific
Fresenius Kabi
accessories,
enables the following functionalities:
Continuous external power supply
Nurse call
Serial link.
An infrared cell is located at the back of the device. It permits information to be exchanged
with the
Link
+
Agilia
: the information can then be transmitted by dedicated communication
cables.
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...