Introduction
1.3 Precautions for use
The
!
symbol visible on the device, indicates that the relevant Instructions For Use should
be read thoroughly.
Volumat Agilia
has been tested in accordance with the standards applicable to
electromagnetic compatibility of the medical devices. Its immunity makes helps to ensure
correct operation. The limitation of the radiation emitted avoids undesirable interference with
other equipment such as EEG, ECG, etc. If a
Volumat Agilia
is placed near devices such as
surgical equipment HF, X-rays, NMR, mobile phones or Wifi points, then minimal distances
between equipment must be respected.
The device must not be used in the presence of inflammable anaesthetic agents due to a risk
of explosion. It should always be used away from all risk areas.
The device could be disturbed by pressure or pressure variations, mechanical shocks, heat
ignition sources, etc. Please contact our Technical Service, if you wish to use devices in
specific conditions. The pump must be used in a horizontal and stable position to work
correctly.
The physiological effects of medicine may be influenced by the characteristics of the device
and the associated tubing (the constitution material is normally listed on the tubing
packaging). Check that the prescriptions are compatible with the characteristics of the
trumpet curves and with the occlusion alarm setting times in relation to the programmed flow
rate.
The device uses a Lithium Ion rechargeable battery. Incorrect handling of a Lithium Ion
battery by a non qualified personnel may cause leakage, heat, smoke, explosion or fire, which
could result in deterioration of performance or failure. This may also damage the protection
device installed in the battery pack, resulting in damage to the equipment or injury to the user.
Fresenius Kabi
cannot, in any circumstances, be held responsible for any medical or other
problem due to incorrect use of the device.
Kindly consult the Instructions For Use for more details.
Use conditions:
Temperature: 5°C to 40°C / 41°F to 104°F.
Humidity: 20% to 90%, no condensation.
Atmospheric pressure: 700 hPa to 1060 hPa.
1.4 Operation safety
As soon as it is in operation, the device ensures continuous surveillance of its functions.
Any internal fault or any procedural anomaly is immediately detected. Nevertheless, abnormal
functioning of the device, without a defined cause, must always be brought to the attention
of the qualified staff in your establishment or our Technical Service.
Volumat Agilia
is equipped with an internal battery that ensures normal functioning during a
mains power cut. In addition, 2 safety fuses protect the electrical mains concerned.
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...