va04.3_001g_Eng.fm
39
Main
tena
nce pr
éven
tive
Date
:
Number
:
Date and number of calibrations.
Select "exit" to return to the "
Test
" menu.
Refer to section 7, "Calibration / Calibration procedure /
Etal 1
pressure", if one of the
values is out of range.
4.4.9 Test 7, Door
This test displays:
the door value in mV and its state, the calibration values in mV, the date and calibration
counter.
Test 7
, press "enter".
The main menu "
Door
" displays.
Scroll through the information, from screen to screen, using
the selection keys.
Conversion
State
Door open calibration value in mV,
Intermediary door calibration value in mV,
Door closed calibration value in mV.
Date
,
Number
:
Calibration date,
Number of calibrations,
Select "exit" to return to the "
Test
" menu.
Refer to section 7, "Calibration/Calibration procedure /
Etal 2
door, if one of the values is
out of range.
Summary of Contents for Volumat Agilia
Page 1: ...0 9 C E S Volumat Agilia Technical Manual ...
Page 6: ...66 TM_Volumat_Agilia_Eng_TDM fm ...
Page 8: ...8 va01_001e_Eng fm Introduction 1 2 Operation diagram ...
Page 12: ...12 va01_001e_Eng fm Introduction ...
Page 26: ...26 va02_001e_Eng fm Description and operation ...
Page 30: ...30 va03 1_001e_Eng fm Description of the menus ...
Page 62: ...62 va04 4_001e_Eng fm Preventive maintenance ...
Page 70: ...70 va05 1_001e_Eng fm Troubleshooting ...
Page 76: ...76 va06 0_001g_Eng fm Intervention procedures ...
Page 82: ...82 va06 0_0011g_Eng fm Intervention procedures ...
Page 86: ...86 va06 0_0021g_Eng fm Intervention procedures ...
Page 94: ...94 va06 0_0040g_Eng fm Fiches d intervention ...
Page 104: ...104 va06 0_0060g_Eng fm Fiches d intervention ...
Page 110: ...110 va06 0_0070g_Eng fm Fiches d intervention ...
Page 134: ...134 va07 1_001g_Eng fm Calibrations ...
Page 136: ...136 va08 0_001d_Eng fm Spare parts catalogue ...