160
3. Connect a calibrator or an actual PT-100 sensor to Input-
1. Calibrate the input using the reference values.
Once this calibration procedure is complete both the logger’s
internal temperature sensor (PT-100) and external PT-100
inputs will be calibrated.
4.14.7.
Calibrating the External PT-100
Sensor Input
PT-100 calibration is a standard 2-point calibration followed by
offset calibration. Once the PT-100 sensor has been selected
in the
Sensor
drop-down menu, click the
Reset Calibration
button. Then proceed with the 2-point calibration procedure.
Before each calibration this step must be performed otherwise
the calibration will not be accurate.
4.14.8.
Saving Calibration Settings
At any time you may manually save the logger’s calibration
settings locally to your PC.
1. Open the logger context menu, select
Calibration > Save
Calibration
and enter the password to access this
feature.
2. The
Save As
dialog will open. Name the calibration file
(with extension .dcf) and save it either in the default
Calibration folder in the DataNet directory or in a folder of
your choice.
4.14.9.
Loading Calibration Settings
You can load a previously saved DataNet .dcf calibration file
at any time to restore a logger’s restore specific calibration
settings.