156
4.14.2.
Calibrating the Data Logger
Figure 51:
Calibration dialog
1. From the Map View, open the Logger context menu and
select
Stop
.
2. Select
Calibration > Calibrate
and enter the password in
the
Password
dialog.
3. Select the sensor you wish to calibrate from the
Sensor
drop-down menu.
4. If calibrating an external sensor, select the inputs to
calibrate. Select
All
or an individual input,
1 to 4
.
5. Click
Setup
to send the sensor setup to the logger.
Note:
The calibration process does not delete the logger’s
setup prior to calibration. Once Calibration is complete the
original logger setup will be restored.