Gas Flow Analyzer
Safety, Error Messages, Troubleshooting, and Support
7
7-11
Problem
Possible Cause
Corrective Action
No airway pressure is
observed.
Wrong flow range (low vs.
high flow) selected.
From the flow screen, ensure
that the range (low vs. high
flow) matches the range
depicted on the side panel of
the
VT PLUS HF
for the port
where the flow is being
applied.
Pressure is negative
when a positive pressure
is applied.
Pressure is being applied to
the "-" connection instead
of the "+" connection.
Ensure that pressure is
applied to the correct
connection.
Ventilator parameters
seem incorrect.
VT PLUS HF
in circuit
backwards
Ensure the inlet side of
the
VT PLUS HF
is connected
closest to the inspiratory
hose from the ventilator. In
bi-directional mode, this
means the inlet is near to the
Y piece.
VT PLUS HF
does not
print.
Wrong printer type selected Select correct printer type on
the
VT PLUS HF
.
Printer not setup for
selected printer language
Review printer's user manual
to determine how to setup
printer t work with the
selected
VT PLUS HF
printer
mode.
Printer not connected
Connect printer.
VT PLUS HF
takes a long
time to establish
communications with PC
software.
This is normal.
If the
VT PLUS HF
does not
eventually communicate with
the PC software, ensure the
cable is correctly connected
and that the PC hardware
can support the
communication port
selected.
Make sure no other device is
trying to use the serial port
on your PC. Close any
software that may be using
the serial port. Make sure
that Windows
recognizes
the serial port in the Control
Panel. Restart the
VT Plus
for Windows
software.
Summary of Contents for VT Plus HF
Page 2: ......
Page 22: ...VT Plus HF Operators Manual 1 10 ...
Page 64: ...VT Plus HF Operators Manual 5 16 ...
Page 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Page 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Page 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Page 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Page 126: ...VT Plus HF Operators Manual 8 36 ...
Page 134: ...VT Plus HF Operators Manual A 8 ...