VT Plus HF
Operators Manual
6-12
5.
Assist Test
The assist test is used to determine the sensitivity of the ventilator in patient assist mode
and to calibrate the full range of assist pressures under which the ventilator can operate.
During an assisted breath, the airway pressure drops during the inspiration due to the
"spontaneous" breath.
Figure 6-1: Normal breath versus an assisted breath. The assist is triggered by a
negative pressure (relative to the end-expiratory pressure) during the expiratory
hold period.
To perform an assist test:
1.
Press the
FULL
key to bring up the
FULL TEST
screen.
2.
Connect the ventilator breathing circuit with the
VT PLUS HF
in the bi-
directional measurement configuration.
Outlet
Inlet
Ventilator
Inspiratory Hose
Expiratory Hose
VT PLUS HF
TEST
LUNG
fig2-1.eps
Figure 6-2 Ventilator Breathing Circuit setup for Bi-directional measurements.
3.
Adjust the ventilator controls so that an effort (negative pressure) at the
lung during the expiration hold period of the ventilator's breath cycle will
trigger the ventilator.
Note:
Use the Synchronized Intermittent Mandatory Ventilation (SIMV) mode
or Intermittent Demand Ventilation (IDV) mode if the ventilator
provides it.
Note:
You will need to test the control setting by applying upward force at the
back of the test lung and observing if the ventilator triggers as desired.
Summary of Contents for VT Plus HF
Page 2: ......
Page 22: ...VT Plus HF Operators Manual 1 10 ...
Page 64: ...VT Plus HF Operators Manual 5 16 ...
Page 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Page 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Page 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Page 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Page 126: ...VT Plus HF Operators Manual 8 36 ...
Page 134: ...VT Plus HF Operators Manual A 8 ...