Gas Flow Analyzer
Safety, Error Messages, Troubleshooting, and Support
7
7-3
4.
Apply 10 lpm of dry air to the high-flow inlet or exhaust.
5.
On the
OXYGEN CALIBRATION
screen select the
ZERO
soft key. Press
OK
to
start 21% O
2
calibration. Allow the full 2-minute period. Do not press skip.
6.
Verify that the O
2
reading is 20.9
±
0.1%.
7.
Apply 10 lpm of dry 100% oxygen to the high-flow inlet or exhaust.
8.
On the
OXYGEN CALIBRATION
screen select the
CAL
soft key. Press
OK
to start
100% O
2
calibration. Allow the full 2-minute period. Do not press
SKIP
.
9.
Verify that the O
2
reading is 100
±
0.1%.
4.
Replacing the Fuses
The
VT PLUS HF's
fuse compartment is located on the side panel of the instrument.
Figure 7-1. The VT PLUS HF Fuse Compartment
To change the fuses:
1.
Turn the
VT PLUS HF
off, and unplug the main power cord.
2.
Using a small, flat-blade screwdriver or similar tool, carefully pry off the cover to the
fuse compartment. The fuse block assembly is attached to the back of the cover.
3.
Carefully lift the old fuses from the fuse slot.
4.
Replace the fuses. (Use two 0.5 A Slo-Blo fuses, Fluke Biomedical P/N 46026.)
5.
Replace the cover on the fuse compartment, and plug in the
VT PLUS HF
.
Warning!
Using other than a recommended fuse can cause fire and/or shock.
Summary of Contents for VT Plus HF
Page 2: ......
Page 22: ...VT Plus HF Operators Manual 1 10 ...
Page 64: ...VT Plus HF Operators Manual 5 16 ...
Page 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Page 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Page 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Page 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Page 126: ...VT Plus HF Operators Manual 8 36 ...
Page 134: ...VT Plus HF Operators Manual A 8 ...