Problem
Possible Cause
Corrective Action
Unit calibration is incorrect
If problem persists, call
Service.
Tidal Volumes are erratic
in bi-directional breath
detect mode
Breath detection threshold
set too low. Noise from
ventilator is causing "false"
breaths.
Set the breath detection
threshold to a higher value.
Tidal volumes different
than expected.
User's expected value was
incorrect.
Review expectations.
Flow baseline drifts,
causing a false non-zero
flow reading when flow was
actually zero.
Zero the flow reading.
Consider using the auto-zero
mode.
VT PLUS HF
is out of
calibration.
If possible, verify calibration
with a 3-liter calibration
syringe. Ensure gas settings
are correct. With the
VT
PLUS HF
in the bi-directional
breath detect mode and the
gas correction mode set to
ATP, stroke syringe such
that flows are near the mid-
range for the flow sensor.
Compare the
VT PLUS HF
tidal volume to calibration
syringe volume. If difference
is out of specification, return
unit for calibration.
Gas settings are incorrect.
Review
VT PLUS HF
gas
settings.
Correction mode is set to
BTPS or STPD instead of
ATP.
Set gas mode to appropriate
setting for your application.
Most applications may
require the ATP mode.
Summary of Contents for VT Plus HF
Page 2: ......
Page 22: ...VT Plus HF Operators Manual 1 10 ...
Page 64: ...VT Plus HF Operators Manual 5 16 ...
Page 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Page 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Page 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Page 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Page 126: ...VT Plus HF Operators Manual 8 36 ...
Page 134: ...VT Plus HF Operators Manual A 8 ...