Gas Flow Analyzer
6
6-9
4.
RT200 Emulation Mode
The
VT PLUS HF
has a function for emulating the RT200 serial port communications.
This is used for some ventilator automated test procedures. To set the RT200 emulation
mode, change the Serial Mode in the
SETUP | SYSTEM
screen:
To access the RT200 Emulation mode, follow these steps:
1.
Press the
MORE/4
key.
2.
Use the
∧
and
∨
soft keys to highlight
RT200 EMULATION MODE
.
3.
Press the soft key below
ENTER
.
OR
1.
Press the
SETUP
key.
2.
Use the
∧
and
∨
soft keys to highlight
SYSTEM
.
3.
Press the soft key below
ENTER.
4.
Use the
∧
and
∨
soft keys to highlight
SERIAL MODE
.
5.
Press the
MODIFY
soft key until
RT200
is displayed.
6.
Press the
BACK
soft key to return to setup menu.
7.
Press the
BACK
soft key again to exit setup and the RT200 screen will
appear.