Uni-directional Flow Mode
Alternatively, the
VT PLUS HF
can be used to measure only inspiratory or expiratory gas
flow as shown in
Figures 2-2
and
2-3
.
Inspiratory Testing
1.
Connect the inspiratory hose to the flow port on the right of the
VT PLUS HF
.
2.
Connect the flow exhaust port on the left to the test lung using standard breathing
hoses.
3.
Ensure that the breath detection mode is set to Uni-directional Inspiratory.
a.
Go into Setup
b.
Choose Settings
c.
Choose Breath Detect and modify as needed
See
Chapter 4, System Setup
for more details.
Outlet
Inlet
Ventilator
Inspiratory Hose
Expiratory Hose
TEST
LUNG
VT PLUS HF
fig2-2.eps
Figure 2-2. Connecting the VT PLUS HF to in either the inspiratory limb of the breathing
circuit.
Expiratory Testing
1.
Connect the expiratory hose to the flow port on the right side of the
VT PLUS HF
.
2.
Connect the flow exhaust port on the left to the ventilator using standard breathing
hoses.
3.
Ensure that the breath detection mode is set to Uni-directional Expiratory.
a
Go into Setup
b
Choose settings
c
Choose Breath Detect and modify as needed
Summary of Contents for VT Plus HF
Page 2: ......
Page 22: ...VT Plus HF Operators Manual 1 10 ...
Page 64: ...VT Plus HF Operators Manual 5 16 ...
Page 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Page 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Page 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Page 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Page 126: ...VT Plus HF Operators Manual 8 36 ...
Page 134: ...VT Plus HF Operators Manual A 8 ...