Gas Flow Analyzer
Specifications
A
A-5
Note:
Fluid pressure may be applied to the positive port; however, fluids should be kept from entering the
pressure port by using a suitable length of connection tubing.
Airway Pressure
Maximum Applied Pressure .........................................................................................20 psi
Operating
Pressure..........................................................................................
±
120 cmH
2
O
Span
Accuracy............................
±
0.75% of reading or
±
0.5 cmH
2
O, whichever is greater
Frequency
Response..................................... > 25 Hz or t
10-90
<40 ms, whichever is greater
Resolution............................................................................................................0.1 cmH
2
O
Sample
rate ...............................................................................................................100 Hz
Note:
Airway pressure is internally tapped off the proximal flow sensor port, which is the port closest to the
exhaust port on the
VT PLUS HF
.
High-Flow Port
Maximum Flow Rate (absolute value).......................................................................500 lpm
Operating
Flow Range ...........................................................................................
±
300 lpm
Accuracy ......................................................................................................
±
2.0% of range
Resolution................................................................................................................0.01 lpm
Frequency
Response..................................... > 25 Hz or t
10-90
<40 ms, whichever is greater
Sample
Rate ..............................................................................................................100 Hz
Dynamic
Resistance ...................................................................... < 2.00 cmH
2
O @ 60 lpm
Low-flow
Dropout........................................................................................................25 lpm
Breath
Detect
Threshold
(user settable) ....................................................................2.0 lpm
Volume
Range ........................................................................................................ > +/- 60 l
Notes:
•
Tidal Volume Ac/- 3% of reading or +/- 10 ml, whichever is greater
•
Volume
Accuracy
tested to 7 liters
•
Flow accuracy is specified for dry air or oxygen
•
Do not use with humidified gases
Low Flow Port
Maximum Flow Rate (absolute value).........................................................................50 lpm
Operating
Flow Range .............................................................................................
±
25 lpm
Accuracy ......................................................................................................
±
1.0% of range
Resolution......................................................0.01 lpm flow > 1 lpm; 0.001 lpm flow < 1 lpm
Frequency
Response..................................... > 25 Hz or t
10-90
<40 ms, whichever is greater
Sample
Rate ..............................................................................................................100 Hz
Dynamic
Resistance .......................................................................... < 2.5 cmH
2
O @ 5 lpm
Low-flow
Dropout.....................................................................................................0.01 lpm
Breath
Detect
Threshold
(user settable) ..................................................................0.50 lpm
Volume
Range .......................................................................................................... > ± 60 l
Notes:
•
Tidal Volume Ac/- 3% of reading or +/- 2 ml, whichever is greater
Summary of Contents for VT Plus HF
Page 2: ......
Page 22: ...VT Plus HF Operators Manual 1 10 ...
Page 64: ...VT Plus HF Operators Manual 5 16 ...
Page 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Page 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Page 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Page 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Page 126: ...VT Plus HF Operators Manual 8 36 ...
Page 134: ...VT Plus HF Operators Manual A 8 ...