VT Plus HF
Operators Manual
4-4
To select
GAS TYPE
, follow these steps:
1.
Use the
∧
and/or
∨
soft keys to highlight
GAS TYPE
.
2.
Press
MODIFY
until the preferred gas type is displayed.
3.
Press
BACK
to return to the settings menu.
Gas Temperature
This setting is the temperature of the gas flowing through the high- or low-flow port.
The
VT PLUS HF
uses this information to improve the accuracy of the flow measurements.
All temperatures are in degrees centigrade.
To enter a new temperature, follow these steps:
1.
Use the
∧
and/or
∨
soft key to highlight the line labeled
GAS
TEMPERATURE
.
2.
Press the
MODIFY
soft key.
3.
Use the numeric keys to enter the temperature in degrees centigrade.
4.
Press the
ENTER
soft key to exit back to the menu options, or to cancel the
change press the
ESCAPE
soft key.
Note that the
VT PLUS HF
accepts temperatures between 0 and 99 °C .
No
decimal values or fractions of a degree are allowed.
Ambient Temperature
This setting is the room temperature air flowing through the high- or low-flow ports.
To change ambient temperature, follow these steps:
1.
Use the
∧
and/or
∨
soft key to highlight the line labeled
AMBIENT TEMP.
2.
Press the
MODIFY
soft key.
3.
Use the numeric keys to enter the temperature in degrees centigrade.
4.
Press the
ENTER
soft key to exit back to the menu options, or to cancel the
change press the
ESCAPE
soft key.
Relative Humidity
This setting is the relative humidity in the air flowing through the high- and low-flow
ports.
To change relative humidity, follow these steps:
1.
Use the
∧
and/or
∨
soft key to highlight the line labeled
REL HUMIDITY.
2.
Press the
MODIFY
soft key.
3.
Use the numeric keys to enter a new value.
4.
Press the
ENTER
soft key to exit back to the menu options, or to cancel the
change, press the
ESCAPE
soft key.
Summary of Contents for VT Plus HF
Page 2: ......
Page 22: ...VT Plus HF Operators Manual 1 10 ...
Page 64: ...VT Plus HF Operators Manual 5 16 ...
Page 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Page 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Page 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Page 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Page 126: ...VT Plus HF Operators Manual 8 36 ...
Page 134: ...VT Plus HF Operators Manual A 8 ...