Chapter 5: Inspect Fiber Endfaces
How to Do the FiberInspector Test
89
Note
To see the buttons for the measurement axes and
core scales and to change the magnification of the
screen, you must first tap
to put the screen in
still mode.
You can use the round, horizontal, and vertical scales to
measure the size of the fiber core and cladding. You can also
measure the size of particles, scratches, and other defects on
the endface.
To show the scales, tap
SCALE ON
(
).
Outer, blue ring: 250
μ
m
Middle, green rings: 120
μ
m and 130
μ
m
Inner, yellow rings: 25
μ
m and 62.5
μ
m (to change the size,
tap
NEXT SCALE
)
To adjust the brightness or contrast of the image, tap
, then
move the bars on the controls. To hide the controls, tap
again.
To show the scales (
), tap
, then tap
SCALE ON
. To change
the size of the measurement ring for the fiber core, tap
NEXT
SCALE
.
Manual
limit selected: To give a
PASS
or
FAIL
grade to the
image, tap
GRADE
. See
.
Test limit selected: To see an analysis of the defects, tap
ANALYZE
. See “Automatic Analysis of Scratches and Defects”
To put the screen in still mode and turn off the probe, tap
.
Tap
to turn on the probe again.
To give the image a grade of
PASS
or
FAIL
, tap
GRADE
(
). If you
set the grade to
FAIL
, then save the image, the ID for the image
is in the list of tests to do again.
Summary of Contents for Versiv
Page 38: ...Versiv Cabling Certification Product Family Users Manual 24...
Page 86: ...Versiv Cabling Certification Product Family Users Manual 72...
Page 202: ...Versiv Cabling Certification Product Family Users Manual 188...
Page 214: ...Versiv Cabling Certification Product Family Users Manual 200...
Page 234: ...Versiv Cabling Certification Product Family Users Manual 220...
Page 244: ...Versiv Cabling Certification Product Family Users Manual 230...
Page 256: ...Versiv Cabling Certification Product Family Users Manual 242...
Page 258: ...Versiv Cabling Certification Product Family Users Manual 244...