9640A
Instruction Manual
3-36
Saving Settings for an Output Function
Settings that apply to the output function include those settings that directly affect the
output signal, but not those that apply to the instrument setup. For example, all of the
settings that contribute to defining a sine output are output function settings. Use the
following procedure to save a set of output function settings:
1.
From the
Save/Recall
screen, select a slot for saving the output function settings.
2.
Press the
Save Mode
soft key. Three new soft labels are displayed:
Save Sine
Mode
,
Save Sweep Mode
, and
Save Mod. Functions
.
3.
Press the appropriate soft key. If the memory slot contains previously saved settings,
the screen prompts for overwrite permission (
Yes
or
No
). Pressing the
Yes
soft key
saves the new settings, and the Type column in the selected slot field displays the
mode to identify the settings as output function settings. Pressing the
No
soft key
aborts the save attempt.
Recalling Settings
Anyone of the 10 saved settings may be recalled at anytime. Use the following
procedure:
1.
From the
Save/Recall
screen, select the slot containing the settings to be recalled.
2.
Press the Recall Settings soft key. The instrument immediately responds to the new
settings.
Creating an RF Output Signal
The Instrument provides three kinds of output signals: sine, modulated, and swept. User
selectable screens, as shown in Figure 3-6, provide control for each of these outputs.
ead15f.bmp
ead94f.eps
ead95f.bmp
Figure 3-6. Control Screens for the RF Output Signal
The remaining sections in this chapter provide the procedures for creating sine,
modulated, and swept output signals. An appropriate screen facsimile and a table
containing a breakdown of the fields accessible on the screen complement each of the
procedures. Procedures for expanded features, such as offset, are presented separately.
Note
Entries displayed in parentheses in the soft labels indicate what will show
in the field after a button is pressed, not what the field currently shows. For
example, if the label indicates
Frequency (Step edit)
, the
Frequency
field
is showing
Cursor edit
.
Note
Many of the data fields in the following procedures include the opportunity
to define measurement units (using the
key). Since the units are often
preferential, it is left to the user to define them. Instructions to do so are not
given in the following procedures.
Summary of Contents for 9640A Series
Page 8: ......
Page 16: ...9640A Instruction Manual viii...
Page 22: ...9640A Instruction Manual 1 2...
Page 40: ...9640A Instruction Manual 2 2...
Page 108: ...9640A Instruction Manual 4A 2...
Page 138: ...9640A Instruction Manual 4C 2...
Page 164: ...9640A Instruction Manual 4D 2...
Page 168: ...9640A Instruction Manual 4D 6...
Page 170: ...9640A Instruction Manual 4E 2...
Page 174: ...9640A Instruction Manual 4E 6...
Page 176: ...9640A Instruction Manual 4F 2...
Page 184: ...9640A Instruction Manual 4F 10...
Page 186: ...9640A Instruction Manual 4G 2...
Page 190: ...9640A Instruction Manual 4G 6...
Page 192: ...9640A Instruction Manual 5 2...
Page 272: ...9640A Instruction Manual 5 82...
Page 274: ...9640A Instruction Manual 6 2...
Page 284: ...9640A Instruction Manual 7 2...
Page 310: ...9640A Instruction Manual 8 2...
Page 316: ...9640A Instruction Manual 8 8...
Page 320: ...9640A Instruction Manual A 4...