![Fluke 9640A Series Instruction Manual Download Page 227](http://html1.mh-extra.com/html/fluke/9640a-series/9640a-series_instruction-manual_2310441227.webp)
Calibration
Performance Test
5
5-37
Table 5-18. Level Accuracy Test (75
Ω
), Low Frequency Test Points
Amplitude
Frequency
Tolerance
1 kHz
±0.06 dB
20 kHz
±0.06 dB
+10 dBm
100 kHz (P
1
) ±0.06
dB
1 kHz
±0.06 dB
20 kHz
±0.06 dB
+7 dBm
100 kHz (P
1
) ±0.06
dB
1 kHz
±0.06 dB
20 kHz
±0.06 dB
-3 dBm
100 kHz (P
1
) ±0.06
dB
1 kHz
±0.06 dB
20 kHz
±0.06 dB
-13 dBm
100 kHz
±0.06 dB
1 kHz
±0.06 dB
20 kHz
±0.06 dB
-23 dBm
100 kHz (P
1
) ±0.06
dB
1 kHz
±0.06 dB
20 kHz
±0.06 dB
-33 dBm
100 kHz (P
1
) ±0.06
dB
1 kHz
±0.06 dB
20 kHz
±0.06 dB
-43 dBm
100 kHz (P
1
) ±0.06
dB
9.
Connect the 75
Ω
Leveling Head to the 75
Ω
power sensor via a precision 75
Ω
Type-N female-to-female adapter. See Figure 5-10 for equipment connections.
Summary of Contents for 9640A Series
Page 8: ......
Page 16: ...9640A Instruction Manual viii...
Page 22: ...9640A Instruction Manual 1 2...
Page 40: ...9640A Instruction Manual 2 2...
Page 108: ...9640A Instruction Manual 4A 2...
Page 138: ...9640A Instruction Manual 4C 2...
Page 164: ...9640A Instruction Manual 4D 2...
Page 168: ...9640A Instruction Manual 4D 6...
Page 170: ...9640A Instruction Manual 4E 2...
Page 174: ...9640A Instruction Manual 4E 6...
Page 176: ...9640A Instruction Manual 4F 2...
Page 184: ...9640A Instruction Manual 4F 10...
Page 186: ...9640A Instruction Manual 4G 2...
Page 190: ...9640A Instruction Manual 4G 6...
Page 192: ...9640A Instruction Manual 5 2...
Page 272: ...9640A Instruction Manual 5 82...
Page 274: ...9640A Instruction Manual 6 2...
Page 284: ...9640A Instruction Manual 7 2...
Page 310: ...9640A Instruction Manual 8 2...
Page 316: ...9640A Instruction Manual 8 8...
Page 320: ...9640A Instruction Manual A 4...