Local Operation
Operating the Instrument
3
3-57
Table 3-20. Sweep-Frequency Fields
ead95f.bmp
Field Range
Units
Start
9.000 Hz to 4.0240000000 GHz
Hz (Hz, kHz, MHz, GHz)
Stop
9.000 Hz to 4.0240000000 GHz
Hz (Hz, kHz, MHz, GHz)
Level
-130.000 dBm to 24 dBm (50
Ω
)
20 dBm max >125.75 MHz
14 dBm max > 1.4084 GHz
-136.000 dBm to 18 dBm (75
Ω
)
14 dBm max >125.75 MHz
8 dBm max > 1.4084 GHz
dBm, Vp-p and Vrms (uV, mV, V),
W (nW, uW, mW, W), dBuV
Linear Step
[1]
0.1 Hz to 4.024GHz
Subject to max steps of 5000000
Hz (Hz, kHz, MHz, GHz), %*, ppm**
and Steps per Sweep
Step Dwell
20 ms to 10 s
Subject to max duration of 100
hrs
s (ms,s)
[1] If a Logarithmic Sweep is selected, the
Linear Step
field is renamed to
Log Step
. The field is now expressed in Units of “Steps per Sweep”
or “Steps per Decade” only. In the latter case Span must be larger than a decade.
* Any entry expressed in % (or converted to %) is subject to a limit of +1000 %.
** Any expressed in ppm (or converted to ppm) is subject to a limit of +10000 ppm.
Measurement Integrity at High Signal Levels
The maximum output level of the Instrument is unusually high (+24 dBm into 50
Ω
and
+18 dBm into 75
Ω
). This power level could damage an RF load, active or passive, or
exceed the maximum-rated level of the load. Measurement integrity can be impaired by
load damage, non-linearity or self-heating of the load.
Measurement Integrity at Low Signal Levels
The Instrument is capable of sourcing very small signal levels (-130 dBm in a 50
Ω
system). At low signal levels, take particular care to eliminate interfering signals from
the measurement. The following notes discuss best interconnection and measurement
practice.
Summary of Contents for 9640A Series
Page 8: ......
Page 16: ...9640A Instruction Manual viii...
Page 22: ...9640A Instruction Manual 1 2...
Page 40: ...9640A Instruction Manual 2 2...
Page 108: ...9640A Instruction Manual 4A 2...
Page 138: ...9640A Instruction Manual 4C 2...
Page 164: ...9640A Instruction Manual 4D 2...
Page 168: ...9640A Instruction Manual 4D 6...
Page 170: ...9640A Instruction Manual 4E 2...
Page 174: ...9640A Instruction Manual 4E 6...
Page 176: ...9640A Instruction Manual 4F 2...
Page 184: ...9640A Instruction Manual 4F 10...
Page 186: ...9640A Instruction Manual 4G 2...
Page 190: ...9640A Instruction Manual 4G 6...
Page 192: ...9640A Instruction Manual 5 2...
Page 272: ...9640A Instruction Manual 5 82...
Page 274: ...9640A Instruction Manual 6 2...
Page 284: ...9640A Instruction Manual 7 2...
Page 310: ...9640A Instruction Manual 8 2...
Page 316: ...9640A Instruction Manual 8 8...
Page 320: ...9640A Instruction Manual A 4...