9640A
Instruction Manual
3-32
Leveling Head
SMA Connector
Torque to 0.45 Nm (4 in-lb)
ead46f.eps
Figure 3-5. Connecting the Leveling Head
Connecting a Leveling Head to a Unit Under Test
The Instrument relies on either a 50
Ω
or a 75
Ω
Leveling Head to maintain the integrity
of the output signal. Both Leveling Heads use N connectors to connect to the input of the
UUT.
Connecting the Leveling Head to a UUT is a straightforward, but critical, process. Before
making the connection, to avoid damage to the instruments involved and to ensure
measurement integrity read and observe the following cautions and warnings,.
W
Caution
To prevent damage to the N connector on the Fluke 9640A-xx
Leveling Heads, use a sacrificial adapter when making frequent
connections or connections to low-quality N connectors.
W
Caution
Reliable and repeatable interconnections are only achieved at
the specified torque setting of 1.00 Nm (9 in-lb). Performance
will be impaired if torque settings are not observed, and
permanent connector damage is likely to result from over-
tightening.
Summary of Contents for 9640A Series
Page 8: ......
Page 16: ...9640A Instruction Manual viii...
Page 22: ...9640A Instruction Manual 1 2...
Page 40: ...9640A Instruction Manual 2 2...
Page 108: ...9640A Instruction Manual 4A 2...
Page 138: ...9640A Instruction Manual 4C 2...
Page 164: ...9640A Instruction Manual 4D 2...
Page 168: ...9640A Instruction Manual 4D 6...
Page 170: ...9640A Instruction Manual 4E 2...
Page 174: ...9640A Instruction Manual 4E 6...
Page 176: ...9640A Instruction Manual 4F 2...
Page 184: ...9640A Instruction Manual 4F 10...
Page 186: ...9640A Instruction Manual 4G 2...
Page 190: ...9640A Instruction Manual 4G 6...
Page 192: ...9640A Instruction Manual 5 2...
Page 272: ...9640A Instruction Manual 5 82...
Page 274: ...9640A Instruction Manual 6 2...
Page 284: ...9640A Instruction Manual 7 2...
Page 310: ...9640A Instruction Manual 8 2...
Page 316: ...9640A Instruction Manual 8 8...
Page 320: ...9640A Instruction Manual A 4...