FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 39 of 131
5.6 Metrological Data Block [8--]
The parameters about Metrologic Registry
are being entered in this section.
[80-] Legal Metrologic Records
[800]
Counter
This counter increases by 1 automatically after entering the programming mode with DIP switch.
This counter cannot be changed manually.
5.7 Diagnostics [9--]
The operations about checking and testing FT-10 FILL can be made here.
[90-] Tests
[900 ] Key Pad testing
In this step every keys ASCII code will be shown on the display as you press the related key. By this
way you can test if all the keys are functional or not. Pressing <
> key will take you to the next
parameter.
[901 ] RS 232C Serial Interface testing
The characters in the alphabet will sequentially be transferred from RS 232C serial interface port by
pressing <
Zero
> key one after another. Received numerical data is seen on display.
[902 ] RS 485 Serial Interface testing
The characters in the alphabet will sequentially be transferred from RS 485 serial interface port by
pressing <
Zero
> key one after another. Received numerical data is seen on display.
[903 ] Parallel Inputs
[ı X Y]
To perform parallel input test, enter the number of parallel input to Y digits via <
Zero
> key. X shows
the logical condition of that input.
[904 ] Parallel Outputs
[o X Y]
To perform parallel output test, enter the number of parallel output to Y digits via <
Zero
> key. To
change the logical condition of that output via <
Tare
> key and X shows the logical condition of that
output.
[905 ] mV Indication
As you press <
Enter
> key the output voltage of the load cell will be shown on the display. This
uncalibrated value is only for test / service purposes.