FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 19 of 131
The HUB connection cabling will be a direct connection as shown below:
Figure 4.7 - HUB connection
The PC connection cabling will be done via cross cable as shown below. IP address blocks and gateway
address of FT-10 FILL and PC should be the same in cross connection.
Figure 4.8
- Cross PC connection
Warning
:
Connect the shield to the reference ground or shield pin of the power connector.
Warning:
Disconnect IndFace1X PC software before starting Modbus-RTU interfacing.
4.3.7 Ethernet Connection (only FT-10 FILL EN)
Ethernet interface is used for data transfer to PC or PLC in the formats shown below.
Usage
Ethernet interface with PC or PLC
Data formats
Continuous, Fast Continuous, Printer Format, BSI Protocol,
Modbus TCP/IP High-Low,
Modbus TCP/IP Low-High
Ethernet
The Ethernet interface operates at 10Mbit, half duplex
Ethernet Connector pin configuration (RJ45)
Pin
Signal
DIR
Description
1
TX+
Out
Differential Ethernet transmit data +
2
TX−
Out
Differential Ethernet transmit data −
3
RX+
In
Differential Ethernet receive data +
6
RX−
In
Differential Ethernet receive data −
4
Not used
Terminated
5
Not used
Terminated
7
Not used
Terminated
8
Not used
Terminated
Shield
Chassis ground
The HUB connection cabling will be a direct connection as shown below:
Figure 4.9 - HUB connection