FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 35 of 131
5.4.3 Scale Block [2--]
[20-] Set Up
The parameters about weighing operation are being entered here.
[200 0] Approved
0 : No
1 : OIML
2 : Hopper and Tank
( * )
( * ) Warning
: Taring, Zero Tracking etc. functions disabled.
[201 0] Increased Indication
0 : by pressing key
1 : Always increased indication
[203 3] Zeroing Range
0 : Disable
1 : ± 2% 2 : ± 20%
3 : ±%50
[204 0] Auto Zero Tracking
AZT automatically readjusts the scale to zero for compensating selected small deviation per second
around centre of zero.
0 : Disable
1 : ± 0,5e
2 : ± 1e
[205 2] Tare
0 : Disable
1 : Multi tare via key
2 : Tare via key if scale is in gross mode
[206 2] Motion Detector
This parameter defines the sensitivity level which will determine what is considered as stable.
0 : ± 0,3e
1 : ± 0,5e
2 : ± 1e
3 : ± 2e
4 : Disable
[21-] Scale Build
The capacity and the resolution of the scale will be defined here.
[212 ] Capacity
Press
key to reach this parameter.
[CAP ] [ XXXXXX ]
Enter the capacity of the scale via
and
keys and confirms the value with pressing
.
[d ] [ XXXXXX ]
Display resolution will be selected by
key and confirmed with
key.
[214 1]
Unit
0 : kg
(Kilogram)
1 : lb
(Libre)
2 : NO
(without unit)