WinFTM File Structure
5
WinFTM File Structure
To ensure correct measurements, a set of specific files needs to be set up. The process
flow for setting up the files needed for measurements, is the same on all FISCHERSCOPE
X-RAY instruments:
Fig. 5-1: Process flow for preparing measurements
The essential content of the files product, measurement application and Def.MA and their
relations among each other is shown in the following figure:
Fig. 5-2: Content of the files Def.MA, measurement application and product
Def.MA:
Definition of the measurement
settings of the application
Product
Measurement results and their presentation
Image of the product
Measurement time and number of measurements
Export masks
Measurement Application
Calibration data (Calibration standard set)
Measurement time during calibration
Number of measurements during calibration
Anode current, aperture
Def.MA
Measurement mode
Type and order of the coatings
Tube voltage
Type and composition of the calibration standards
Treatment of interfering spectra
Special evaluation methods