Mass per unit area and geometric coating thickness
Calibration standards from FISCHER
94
FISCHERSCOPE
®
X-RAY
Elements, Heräus GmbH/Hanau, Reference 5C 3.81\N Ko).
However, they can be changed by the user.
13.2 Calibration standards from FISCHER
The density information of all calibration standards are computed
according to the values of the Heräus table, and are stated
accordingly on the certificates. Thus, the coating information of the
calibration standards have been determined according to the mass
per unit area method.
No data have been obtained using the cross-cut method.
By default, WinFTM uses the density values of the Heräus table.
The calibration lab of FISCHER has been accredited by the
Deutsche Akkreditierungsrat [German Accrediting Council] for the
measurement unit “mass per area unit”.
13.3 Setting the density in the Def.MA
You can manually set the density of an element in the Def.MA.
Please observe the notes of the previous sub-chapters.
Dialog window “Def.MA”. The current density setting for each
element is shown in the left window.
To open the dialog window “Def.MA”:
The menu command “Calibrate > Def.MA new > Thickness mode”
opens the dialog window “Def.MA” in the thickness mode.
The thickness mode is the most frequent mode selection.
Summary of Contents for FISCHERSCOPE X-RAY 4000 Series
Page 18: ...18 FISCHERSCOPE X RAY Components...
Page 24: ...24 FISCHERSCOPE X RAY Manual Measurements Deleting Measurement Readings...
Page 28: ...28 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 44: ...44 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 122: ...122 FISCHERSCOPE X RAY Calibration...
Page 140: ...140 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 167: ...WinFTM 167...