40
FISCHERSCOPE
®
X-RAY
User Interface of the WinFTM Software
The Spectrum Window
You can change the display of the X axis to show the energies directly.
To toggle the display of the X axis between energies and channels
►
Select in the
Spectrum
window
Scaling > X-Axis energy
.
To display the spectrum as colored area or colored line
►
In the window
Spectrum
click
. With this button you can toggle between both repre-
sentations.
To display the spectral lines of all elements of the current product
►
Select in the
Spectrum
window
View > Elements of Product
.
6.8.1
Saving a Spectrum
You can use a saved spectrum to re-evaluate it with a different measurement application,
without the need of performing the measurement again.
To save a a single spectrum
You can save a single spectrum in the WinFTM main window or in the
Spectrum
window.
1. Depending on the window that is currently opened, do one of the following:
The
Spectrum
window is opened: Select
Spectrum > Save as…
.
The WinFTM main window is opened: Select
General > Save Spectrum…
.
The window
Save as
appears. A default file path is specified, which contains the
current high voltage and the current primary filter, if available. Using this file path
makes it easy to retrieve the spectrum later.
2. Enter a file name and click
Save
.
3. The file will be saved with the extension
*.spk
or
*.spe
.
To save the spectrum after every measurement automatically
1. In the WinFTM main window select
Store Spectrum Automatically
. If the automatic
storing is activated, a disk symbol appears on the
Start
button.
Repeat the step to deactivate the automatic storing.
6.8.2
Loading a spectrum
You can load a saved spectrum for evaluation purposes at any time. You can also load
several spectra to compare them.
1. In the window
Spectrum
click .
The window
Open
appears.
2. Select the spectrum and click
Open
.
If a spectrum is already displayed in the spectrum window, then the newly loaded spec-
trum will be displayed in a different color.
Summary of Contents for FISCHERSCOPE X-RAY 4000 Series
Page 18: ...18 FISCHERSCOPE X RAY Components...
Page 24: ...24 FISCHERSCOPE X RAY Manual Measurements Deleting Measurement Readings...
Page 28: ...28 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 44: ...44 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 122: ...122 FISCHERSCOPE X RAY Calibration...
Page 140: ...140 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 167: ...WinFTM 167...