FISCHERSCOPE
®
X-RAY
21
Manual Measurements
4
Manual Measurements
Usually the instrument is integrated into production lines. Measurements are performed
automatically, controlled by a superordinate system. Measurement results are sent to the
superordinate system.
But you can also measure manually. This section describes how to perform measure-
ments manually.
To perform measurements you need a product. A product is a file, which contains all
settings for a specific measurement task.
When the instrument is delivered by FISCHER, products that cover your measurement
tasks are usually available. With these products you can perform measurements
immediately.
If you need new measurement tasks, you can request products from FISCHER. The
products are then send to you by email or on a data carrier. Before you can use the new
products, you need to import and calibrate them.
This section describes how to perform measurements.
For further information see the following chapters:
Before you start
The instrument is switched on (see Chap. 3.1 on Page 19).
An appropriate product file is present.
Importing a product
Chap. 7.3 on Page 46
Performing a calibration
Chap. 15 on Page 117
Summary of Contents for FISCHERSCOPE X-RAY 4000 Series
Page 18: ...18 FISCHERSCOPE X RAY Components...
Page 24: ...24 FISCHERSCOPE X RAY Manual Measurements Deleting Measurement Readings...
Page 28: ...28 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 44: ...44 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 122: ...122 FISCHERSCOPE X RAY Calibration...
Page 140: ...140 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 167: ...WinFTM 167...