FISCHERSCOPE
®
X-RAY
125
Export Settings
Measurement Data Export
Table 16-2: Explanation of the controls of the Exporting data window
Control
Explanation
Online-Export ON
The measurement data is exported continuously when
measurement data is acquired by the instrument.
The reading is exported directly after a measurement is made.
The block statistics are exported upon block closure.
With product selection, the header is exported.
Template and No.
The name and the number of the export template are displayed if
an export template is assigned to the current Product.
►
To assign an export template, click
Select…
.
Deletes the assignment of an export template to the product.
Export destination
append to file
Data is successively written to a file.
Complete file-
handshake
(NET_EXPT.END)
WinFTM waits until the handshake file
NET_EXPT.END
is
deleted. This control box should be enabled only, if the technology
of the target computer is capable of handling it.
overwrite file
The export file contains only the last exported values, i. e. the old
data will be overwritten.
Destination file
Displays the destination file, if already assigned.
►
To assign a destination file for the data export, click
Select…
.
Filename variable
By clicking the list box you can select a rule, how the file name of
the exported data will be created.
send to serial
port
Data is sent directly to the RS232 interface.
The parameters of the RS232 interface and of the WinFTM
software of the receiving instrument must be the same. Otherwise
the receiving instrument will not be able to read the data exported
via the RS232 interface.
►
To set the parameters of the RS232 interface select
Evaluation > Export > Define export interface…
.
to OLE-
Automation
WinFTM can be remote controlled via OLE and can transfer data
via OLE. A Visual-Basic (Version 6) program example is included
on the WinFTM installation CD in the directory
Service\OLE_Import_Sample_VisualBasic_6.0\AutomationDemo
.
This sample program can be used as a template for the customers
own remote control software.
Summary of Contents for FISCHERSCOPE X-RAY 4000 Series
Page 18: ...18 FISCHERSCOPE X RAY Components...
Page 24: ...24 FISCHERSCOPE X RAY Manual Measurements Deleting Measurement Readings...
Page 28: ...28 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 44: ...44 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 122: ...122 FISCHERSCOPE X RAY Calibration...
Page 140: ...140 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 167: ...WinFTM 167...