34
FISCHERSCOPE
®
X-RAY
User Interface of the WinFTM Software
Measurement Value Field
If you have set up tolerance limits, the measurement value is displayed with colored back-
ground. If the measurement value is within the tolerance limits, the measurement value is
displayed on green background, otherwise on red background.
6.4.3
SPC Chart Presentation
In this display mode the SPC chart (Statistical Process Control chart) is displayed in the
statistics field. The measurement values are displayed in the list presentation in the
measurement value field.
Within the statistical process control chart you can identify trends in automated measuring
procedures with large lot sizes. Thus, you can easily determine changes with regard to
coating thickness and other process parameters. Furthermore, the SPC chart presentation
simplifies the monitoring of the measurement devices.
Fig. 6-3: SPC chart presentation, the upper image shows the mean value over time,
the lower image shows the standard deviation over time
6.4.4
Changing the Measurement Value Presentation
To display or hide result channels
WinFTM BASIC/SUPER/PDM can process up to 24 result channels (coating thickness
values or analytical parameters). WinFTM LIGHT can process up to four coating thickness
values or up to five analytical parameters. The result channels are defined within the
DefMA.
Summary of Contents for FISCHERSCOPE X-RAY 4000 Series
Page 18: ...18 FISCHERSCOPE X RAY Components...
Page 24: ...24 FISCHERSCOPE X RAY Manual Measurements Deleting Measurement Readings...
Page 28: ...28 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 44: ...44 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 122: ...122 FISCHERSCOPE X RAY Calibration...
Page 140: ...140 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 167: ...WinFTM 167...