WinFTM
®
165
Comparative Overview with/without WinFTM PDM
WinFTM SUPER: For the Experienced X-RAY User
4
WinFTM SUPER: For the Experienced X-RAY User
For both WinFTM versions BASIC and LIGHT the extension SUPER is available. With
WinFTM SUPER you can change the following parameters:
Measurement mode
Tube voltage used
Type and order of the coatings
Type and composition of the substrate material of the specimen and of the calibration
standards
Treatment of interfering spectra
Special evaluation methods
These parameters are stored in the Def.MA file (
Def
inition of the
M
easurement Settings
of the
A
pplication).
The usage of the SUPER software requires deep knowledge of the physics of x-ray fluo-
rescence and the functionality of the instrument.
Alternatively, you can request a new measurement application from FISCHER. FISCHER
offers a professional service for the creation of customer’s Def.MA.
In addition, the SUPER software features the following functions:
Function
Description
Convenient entering of
elements
Convenient entering of elements, e.g., in the Def.MA using
the window
Periodic System
.
Pool of Def.MA files
You can establish a pool of Def.MA files. Several users can
access to this pool.
Summary of Contents for FISCHERSCOPE X-RAY 4000 Series
Page 18: ...18 FISCHERSCOPE X RAY Components...
Page 24: ...24 FISCHERSCOPE X RAY Manual Measurements Deleting Measurement Readings...
Page 28: ...28 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 44: ...44 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 122: ...122 FISCHERSCOPE X RAY Calibration...
Page 140: ...140 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 167: ...WinFTM 167...