FISCHERSCOPE
®
X-RAY 4000 Series
143
General Specification
Intended use
Energy dispersive x-ray fluorescence measuring system (EDXRF) to analyze and
measure coatings and layer systems continuously in production processes.
Element range
Instruments with proportional counter tube: Chlorine (17) to Uranium (92)
Instruments with PIN diode or silicon drift detector: Aluminium (13) to Uranium (92)
Features
Calibration
Fully automated, can be carried out simply and quickly
Process monitoring
Warning flashlight for control of exceeding and underscoring limits helps to control the
manufacturing process
Remote control
Via programmed tasks or via various data interfaces
Date Interface
RS232, Command files, OLE automation, TCP/IP, PROFIBUS and PROFINET via
OPC
Maximum width of the sample
950 mm (37 in)
X-Ray Source
X-ray tube
Tungsten tube with glass window or microfocus tungsten tube with berylium window
High voltage
Three steps:
Instruments with proportional counter tube: 30 kV, 40 kV, 50 kV
Instruments with PIN diode or silicon-drift-detector: 10 kV, 30 kV, 50 kV
Aperture
(Collimator)
2x changeable: Ø 0.3 mm (12 mils) and 4 mm x 0.12 mm (157 mils x 5 mils),
or 4x changeable according to customers' requirements
Primary filter
Instruments with proportional counter tube: fix or 3x changeable
Instruments with PIN diode or silicon-drift-detector: fix or 6x changeable
X-Ray Detection
X-ray detector
Proportional counter tube, peltier-cooled Silicon PIN diode or silicon drift detector
Resolution
(fwhm for Mn-K
α
)
Instruments with proportional counter tube: approx. 900 eV
Instruments with PIN diode: up to
≤
180 eV
Instruments with silicon-drift-detector: up to
≤
140 eV
Signal processor
Instruments with proportional counter tube: analog signal processing
Instruments with PIN diode or silicon-drift-detector: digital pulse processor
Distance compensation
± 3 mm (118 mils) for measurements on glass substrates, with specific applications on
glass substrates up to ± 5 mm (197 mils)
Measuring distance
12.5 mm (0.5 in) from the housing, other distances on request
Electrical data
Power supply
AC 115 V or AC 230 V 50 / 60 Hz
Power consumption
max. 120 W (without evaluation PC)
Protection class
IP50
Dimensions
4100
4200
4300
Weight
120 kg
140 kg
160 kg
Dimensions W x D x H
596 x 1615 x 308 mm
(23 x 64 x 12 in)
236 x 350 x 370 mm
(9 x 14 x 15 in)
569 x 2015 x 308 mm
(22 x 79 x 12 in)
Travel
240 mm (9 in)
620 mm (24 in)
1000 mm (39 in)
Positioning accuracy
≤
0.02 mm (0.8 mils)
≤
0.02 mm (0.8 mils)
≤
0.02 mm (0.8 mils)
Summary of Contents for FISCHERSCOPE X-RAY 4000 Series
Page 18: ...18 FISCHERSCOPE X RAY Components...
Page 24: ...24 FISCHERSCOPE X RAY Manual Measurements Deleting Measurement Readings...
Page 28: ...28 FISCHERSCOPE X RAY WinFTM File Structure Product...
Page 44: ...44 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Page 122: ...122 FISCHERSCOPE X RAY Calibration...
Page 140: ...140 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Page 167: ...WinFTM 167...