ALINTE
4
PAGE
I
NSTALLATION
G
UIDE
5-11
The plate will eject when the test is complete.
10.
Process the plate to reveal the three sets of test strips:
The first two sets are for 96 resolution.
The third set is for 48 resolution.
Table 5.1 Beam focus check
AUTO FOCUS
MACRO:
RUN TEST STRIPS
Focus 96V
Focus 96H
Focus 48V
Set Defaults?
Yes
N/A
N/A
Scan Start Posn (mm)
10
N/A
N/A
Strip Width (mm)
8
8
12
Start Value
-900
-900
2500
Step Value
50
50
200
Number of Steps
13
13
11
Resolution
96 I/mm
96 I/mm
48 I/mm
Pattern Type
Vert
Hor
Vert
Lines
2on 2off
2on 2off
1on 1off
Start Scan
Yes
Runs as Auto
Runs as Auto