![Fenwal 4R5707 Operator'S Manual Download Page 434](http://html.mh-extra.com/html/fenwal/4r5707/4r5707_operators-manual_553093434.webp)
A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
A
PPENDIX
A-6
07-19-01-518 – January 2009
The table below indicates the maximum blood donation volumes
cleared by the FDA for collection of red blood cells and plasma
using the A
LYX
RBC /Plasma procedure.
Donor
Weight
Donor
Hematocrit*
[%]
Donor
Hemoglobin
[g/dL]
Total
Absolute
RBC
Volume
[mL]
Maximum
Plasma
Volume
[mL]
Male or
Female
MIN
MAX
MIN
MAX
110 - 129 lbs.
38
55
12.5
18.3
200
450
130 – 174 lbs.
38
55
12.5
18.3
200
550
175 lbs and
over
38
50
12.5
16.6
200
650
175 lbs and
over
51
55
16.7
18.3
200
550
*Pre-donation hematocrit may be determined using a 5 minute spun hematocrit or
equivalent quantitative method.